Introduction

The device acquires functions that including synchronous sampling of conventional 12-lead ECG data and 3-lead vector ECG data, various forms of analysis and diagnosis, report output and other functions. As an integrated ECG analysis system, the device can reflect user's heart movement timely and accurately. It is applicable for ECG examination in medical and health institutions, providing accurate and reliable basis for diagnosis.


Features
1.Compact and portable design for Recorder, support both Wilson and Frank lead system, with lead-off indicator..
2.Adopt real-time sampling, display and storage mode, with functions such as QRS waveform analyzing in real-time, HR calculation and real-time printing, as well as multiple filter methods to avoid from baseline drift and interference.
3. Advanced ECG analysis algorithm ensures accurate auto-analyzing points and reliable auto-recognizing sort of arrhythmia, which makes the analyzing report and conclusion more valuable.
4. For clear display, ECG waveform uses anti-aliasing technology; smart adjustment function of lead position solves the problem of overlapping waveform caused by baseline drift or high amplitude waveform.
5. High sampling frequency (1000 Hz) ensures correct ECG waveform without distortion. High resolution in display and printout guarantees the details of ECG waveform.
6. With dozens of sync and continuous display/print modes, including 6x2, 3x4, 3x4+2, etc. Auxiliary tool, such as superposed waveform analysis, electronic scale and editable terminology database, etc., brings much more convenient for diagnosis and clinical application.
7. Multiple analysis function, including: conventional 12-lead ECG, Frequency Cardiogram (FCG),High Frequency ECG, QT internal dispersion analysis, Heart Rate Variability (HRV) analysis, myocardium ischemia (STLE) analysis, Heart rate turbulence (HRT) analysis, Pacing ECG analysis, Vector Cardiogram (VCG),Time Vector Cardiogram (TVCG) and SAECG, etc.
8. The "shortcut print" function can preview or print the whole analysis reports, and save them in PC with multimedia format (BMP, PNG, PDF or JPG format) according to factual need
9. Support calling external procedures to obtain patient information of unknown origin, exporting HL7 aECG standard format ECG data, printing electronic report and sending notice to external procedure, convenience for systems docking with hospital HIS, PACS and others.


Performance
HR display range: 30 bpm~300 bpm, error: 30 bpm~100 bpm, ±1b pm; 100 bpm~ 300 bpm, ±1%.
Sampling frequency of ECG signal input: 32 k Hz
Sampling frequency of waveform data processing: 1 kHz
Sampling accuracy: 24 bits
Lead: Standard 12 leads
Dynamic range: ±5 mV
Gain control: continuous variation is available
Gain switch: When the gain is changing, there is indication for the recording output.
Gain stability: After startup for 1 minute, the gain variation per minute shall be no more than 0.33%. The total variation within 1 hour of the constant gain at each level shall be no more than ±3%.
Time base accuracy: within 0.2 s~2.0 s, the permitted error shall be no more than ±5%.
Channel width: the display width of each channel shall be no less than 40mm.
Track width: <1 mm
System error: ±5%
Lead weight factor: accuracy within ±5%
Calibration voltage: 1 mV, deviation within ±5%
Input impedance: >2.5 MΩ; >50 MΩ (10 Hz)
Patient-electrode DC current: <1 μA
System noise: ≤15 μVp-p
CMRR: >105 dB
Amplitude-frequency characteristic:
Rated input amplitude Input frequency and waveform Relative output response
1.0 0.67 Hz ~ 40 Hz, sinusoidal ±10%a
0.5 40 Hz ~ 100 Hz, sinusoidal +10%, -30%a
0.25 100 Hz ~ 150 Hz, sinusoidal +10%, -30%a
0.5 150 Hz ~ 500 Hz, sinusoidal +10%, -100%a
1.5 ≤1 Hz ,200 ms, triangular +0%, -10%b
a Output relative to 10 Hz
b Output relative to 200 ms
Gain setting and accuracy: three levels: 5, 10, 20 mm/mV, accuracy: ±5%
Time base: 12.5 mm/s, 25 mm/s, 50 mm/s
Safety type: Class II, Type CF with defibrillation protection


Accessories
Ground wire 1pc
Patch cord 2pcs
Data transmission line 1pc
Lead cable 1pc
Limb electrode clamp 1 set (4pcs)
Chest electrode 1 set (6pcs)
User manual 1pc
lead wire clip 1pc(2pcs)
software 1 pc


Physical Characteristic
Operational environment
a) Environment temperature: 5℃~ 40℃
b) Relative humidity: 25%~ 95% (no condensation)
c) Power supply: USB DC 5V
d) Atmospheric pressure: 700hPa~1060hPa
Storage environment
a) Environment temperature: -40℃~+55℃
b) Relative humidity: ≤95%HR
c) Atmospheric pressure: 500hPa~1060hPa

Certification
CFDA certificate ,CMC certificate and CE certificate.

Pogledajte više
220.0$

CONTEC8000G PC Based ECG Workstation 12 Lead Resting System Recorder USB PC SW

Dostava uredi Brazil

Model: CONTEC8000G

CONTEC8000G
1
Na lageru: 638
Naručite odmah
Dodaj u korpu
Objavljeno: 25.11.2024 19:01 Na lageru: 638

Oglasi iz iste kategorije